Part number: LPV448.99.00001
The contents of the box
TOC Instrument
Sample tray with 65 positions for 40 mL vials
Glass sample vials 40 mL (100 pcs)
USB 2.0 cable AB
Set of tubing PTFE blue-stripe 1/8″ OD length 2 m including nut & ferrule
Reagent container with screwcap for UPW, Acid and Waste
Power cord
USB storage device with software, test results and Full User Manual
Printed User Manual
Specifications
Power supply:Analyzer: 100 – 240 V ac, 50/60 Hz, 16 A, with protective ground
PC: 100 – 240 V ac, 50/60 Hz, 1.6 A, with protective ground
Monitor: 100 – 240 V ac, 50/60 Hz, 1.6 A, with protective earthing
Operating conditions:20 – 30 °C; 20 – 80% relative humidity (does not cause condensation)
Gas consumption:150 mL/min*
*Sample preparation for NPOC determination requires an additional 300-350 mL/min.
Box contents:TOC Instrument
Sample tray with 65 positions for 40 mL vials
Glass sample vials 40 mL (100 pcs)
USB 2.0 cable AB
Set of tubing PTFE blue-stripe 1/8″ OD length 2 m including nut & ferrule
Reagent container with screwcap for UPW, Acid and Waste
Power cord
USB storage device with software, test results and Full User Manual
Printed User Manual
Dimensions:440mm x 380mm x 700mm (H x W x D)
Measuring range:TC, TIC, NPOC: 0 – 30000 mg/L
Degree of pollution:2
Weight:Approx. 31 kg
Detection limit:TC, TOC, NPOC: 50 µg/L
Measurement method:NDIR (non-dispersive infrared detection)
Oxidation method:Catalytic combustion at 680 °C
Model:QP1680-TOC
Regulations and Standards:TOC / NPOC: ASTM D7573, EN 1484, EPA 415.1, EPA 9060A, ISO 8245, SM 5310B, NEN-ISO 20236
Number of sample positions:65
Parameter:heel
Power:Analyzer: 750 W max.
PC: 90 W max.
Monitor: 100 W max.
Repeatability:Up to 10 mg/L TC, TIC, NPOC, TN: < 5%
> 10 mg/L TC, TIC, NPOC, TN: < 2%
Gas specifications:Oxygen or synthetic air: minimum 99.998% (4.8) at 3 – 10 bar
Ambient temperature:-20 – 60 °C
Temperature:Max. temperature of the oven 1050 °C (depending on the configuration)
Analysis time:Approx. 3 minutes
Sample volume:10 – 1000 µL