Part number: LPV448.99.01001
The contents of the box
TOC/TNb Instrument
Sample tray with 65 positions for 40 mL vials
Glass sample vials 40 mL (100 pcs)
USB 2.0 cable AB
Set of tubing PTFE blue-stripe 1/8″ OD length 2 m including nut & ferrule
Reagent container with screwcap for UPW, Acid and Waste
Power cord
USB storage device with software, test results and Full User Manual
Printed User Manual
Specifications
Power supply:Analyzer: 100 – 240 V ac, 50/60 Hz, 16 A, with protective ground
PC: 100 – 240 V ac, 50/60 Hz, 1.6 A, with protective ground
Monitor: 100 – 240 V ac, 50/60 Hz, 1.6 A, with protective earthing
Operating conditions:20 – 30 °C; 20 – 80% relative humidity (does not cause condensation)
Gas consumption:200 mL/min*
*Sample preparation for NPOC determination requires an additional 300-350 mL/min.
Box Contents:TOC/TNb Instrument
Sample tray with 65 positions for 40 mL vials
Glass sample vials 40 mL (100 pcs)
USB 2.0 cable AB
Set of tubing PTFE blue-stripe 1/8″ OD length 2 m including nut & ferrule
Reagent container with screwcap for UPW, Acid and Waste
Power cord
USB storage device with software, test results and Full User Manual
Printed User Manual
Dimensions:440mm x 380mm x 700mm (H x W x D)
Measuring range:TC, TIC, NPOC, TN b : 0 – 30000 mg/L
Degree of pollution:2
Weight:Approx. 32 kg
Detection limit:TC, TIC, NPOC: 50 µg/L
TN b : 20 µg/L
Measurement method:TOC: NDIR (non-dispersive infrared detection)
TN: Chemiluminescence
Oxidation method:Catalytic combustion at 720 °C
Model:QP1680-TOC/TN b
Regulations and Standards:TOC / NPOC: ASTM D7573, EN 1484, EPA 415.1, EPA 9060A, ISO 8245, SM 5310B, NEN-ISO 20236
TN b : ASTM D8083, EN 12260, ISO 11905-2, NEN-ISO 20236
Number of sample positions:65
Parameter:OCD, TN b
Power:Analyzer: 750 W max.
PC: 90 W max.
Monitor: 100 W max.
Repeatability:Up to 10 mg/L TC, TIC, NPOC, TN: < 5%
> 10 mg/L TC, TIC, NPOC, TN: < 2%
Gas specifications:Oxygen or synthetic air: minimum 99.998% (4.8) at 3 – 10 bar
Ambient temperature:-20 – 60 °C
Temperature:Max. temperature of the oven 1050 °C (depending on the configuration)
Analysis time:Approx. 4 minutes
Sample volume:10 – 1000 µL