Part number: LPV441.99.00011
Your water testing requirements, all in one spectrophotometer
The DR6000 instrument offers the most pre-programmed test methods, including high-speed wavelength scans in the UV and visible spectrum.
Designed for high volume and exceptional precision
A carousel sample changer allows up to seven sequential measurements. The Sipper module, an instrument-controlled sample delivery system, increases precision through constant optical characteristics.
Advanced quality assurance at your fingertips.
The DR6000 instrument comes with integrated QA software for programming, documenting and interpreting all necessary quality measurements.
Guided procedures for eliminating false readings
When used with LCK cuvette tests, the DR6000 provides the precise results you need to guide you step-by-step through testing procedures. Through LCK cuvette tests, the instrument averages 10 readings and eliminates outliers, meaning scratched, damaged or dirty glass containers are not a problem.
Avoid errors automatically
RFID* technology automatically updates program calibration factors when you place the LCK reagent box next to the DR6000. The tool identifies the expiration dates of chemicals by means of a barcode on the vials, and detects the factors of the chemical coefficients, to avoid errors that intervene in the variations from one batch to another in the chemicals.
*RFID technology is available in all EU countries plus Norway, Switzerland, Serbia, Macedonia, Turkey, Russia and others. For other countries, consult your local Hach contact.
The contents of the box
1 x Power Cord (country specific: EU, CH, UK)
1x Universal-Adapter
1x Dust Cover
1x Printed multilingual basic user manual (bg, cz, da, en, es, fr, gr, hr, hu, it, nl, pl, pt, ro, ru, sl, sl, sv, tr)
Specifications
Accessory:Application software: LZV935 for drinking water analysis, LZV936 for brewery analysis, LZV937 for enzymatic food analysis, LZV902.99.00001 – carousel holder for enzymatic food testing
Photometric accuracy:1 % at 0.50 – 2.0 Abs
5 mAbs at 0.0 – 0.5 Abs
Wavelength accuracy:± 1 nm
Wavelength selection:Automatic
Waterproof housing (IP) :IP20 with the cover closed
Bowl compatibility:Rectangular: 10, 20, 30, 50 mm, 1 inch; cylindrical: 13 mm, 1 inch
Optional 100 mm rectangular cell with additional adapter
Bowl compatibility:Optional 100 mm rectangular cell with additional adapter
Storage conditions:-25 °C – 60 °C, relative humidity max. 80% (without condensation)
Operating conditions:10 – 40 °C, relative humidity max. 80% (without condensation)
Box contents:1 x Power Cord (country specific: EU, CH, UK)
1x Universal-Adapter
1x Dust Cover
1x Printed multilingual basic user manual (bg, cz, da, en, es, fr, gr, hr, hu, it, nl, pl, pt, ro, ru, sl, sl, sv, tr)
Weight:11 kg
IBR+:Automatic test recognition, batch control and expiry date verification
Data logger:5000 measured values (result, date, time, sample ID, operator ID)
Interfaces (options):USB type A (2), USB type B, Ethernet, RFID module
Photometric measurement range:± 3 Abs (wavelength range between 340 and 900 nm)
Wavelength range:190 – 1100 nm
Spectral bandwidth:2 nm
User interface language:Bulgarian, Chinese, Croatian, Czech, Danish,
Dutch, English, French, German,
Greek, Hungarian, Italian, Japanese,
Polish, Portuguese,
Russian, Slovak, Slovenian,
Spanish, Swedish, Turkish
Manual languages:German, English, French, Italian, Spanish, Portuguese (PT), Czech, Danish, Dutch, Hungarian, Polish, Romanian, Russian, Slovenian, Slovak, Swedish, Turkish, Greek, Croatian, Bulgarian, Slovak
Photometric linearity:1 % at >2 Abs with neutral glass at 546 nm
0.005 – 2 Abs
LINK2SC:Data exchange with the SC 1000 controller
Scatter light:KI solution at 220 nm <3.3 Abs
Mode of operation:Transmittance (%), absorbance, concentration, scan
Monitor:7″ TFT WVGA color touch screen
Pre-programmed methods:> 240
Printer compatibility:Compatible with most office inkjet printers
User programs:two hundred
Quality assurance:Programming and documentation function of the quality assurance procedure, with indication of success/failure
Network connection:50 – 60 Hz
Power cable, 100 – 240 V, 50 – 60 Hz
Wavelength reproducibility:< 0.1 nm
Wavelength resolution:0.1 nm
Optical system:The reference beam, spectral
Stray lights:< 0.0005
Lamp source:Tungsten (VIS), deuterium lamp (UV)
Specific technology:RFID for easy updating of methods, sample identification data and certificates of analysis
Scan speed:900 nm/min (in 1 nm steps)